Oil-in-Water Dual-Channel Handheld Fluorometer

 

PAHs (Polycyclic Aromatic Hydrocarbons) are one of the most widespread organic pollutants. In addition to their presence in fossil fuels, they are also formed by incomplete combustion of carbon-containing fuels such as wood, coal, diesel, fat, tobacco, and incense. Due to their physical properties, PAHs are widely used in many industrial applications, such as lubricating oil, hydraulic oil, and electro-hydraulic control fluid. Their applications also could produce man-made contamination in the water systems due to leakage of the oils.

➥ Fluorescence Detection:
The aromatic fraction of PAHs can be excited with UV light (360nm for crude oil) or deep-UV light (280nm for refined oil) to emit fluorescent light. The fluorescence intensity is linear with the concentration of the oil. This technology is ideal for monitoring leaking lube oil in the cooling systems or power plants, or leaking crude oil in ocean/lake environment.
Our handheld dual-channel fluorometer is designed for oil and/or fuel detection and monitoring. Due to the low interference from other substances, we have proven that this technology can detect sub-mg/L (<1-ppm) level of oil/fuel contamination in water samples from the environment. And due to its high portability, it can be used anywhere in the field to conduct environmental inspection for potential oil contaminations.

 

➥ Fluorometer Specifications:
● Uses proprietary quartz mini-tube for easy sample collection and high sensitivity.
● Rapid measuring (5 seconds reading) and user-friendly interface.
● Sensitive and wide measurement range.
  Typical minimum detection limits:
  -- Refined Oil: 1-5, naphthalenedisulfonic acid disodium salt (10 – 10,000ppb).
  -- 5W30 Motor oil (1 – 1,000ppm). Gasoline (2 – 5,000ppm).
  -- Crude Oil: Crude oil (0.1 – 1,000ppm).
● Simple touch screen calibration. No repeated calibrations needed.
● Powered by 9V (alkaline or lithium) battery*, or 9VDC power adaptor.
● Portable for field operation, and store up to 2x80 data points for computer analysis.

 

➥ Field kit includes the following items:
*Dual-Channel Handheld Fluorometer for Refined Oil and Crude Oil
*Disposable 500-μL Transfer Pipette: 100 pcs
*Re-useable Quartz Mini Tube: 10 pcs

 

➠ See the detailed information: Oil-in-Water Dual-Channel Handheld Fluorometer